Huestis Medical remanufactured OEC 9600 C-arms offer excellent imaging results at a remarkable price.  Even more remarkable is our ability to deliver performance and safety features guaranteed to meet or exceed original manufacturer’s specification.

Thoroughly remanufactured, our OEC 9600 C-arms are offered with new and upgraded imaging components.  We disassemble, mechanically restore, refinish, calibrate and pre-stage each C-arm prior to delivery.

Process / Features

  • Mechanical repairs/cosmetic maintenance
  • Computer controlled ABS, mA, kVp and camera gain
  • All defective parts and covers are repaired or replaced
  • C-arm is completely disassembled, disinfected and repainted
  • “C” is removed from main frame assembly
  • Horizontal arm removed, inspected and repaired as necessary
  • Vertical column removed and bearings removed, inspected and replaced
  • X-ray tube removed and inspected for artifacts and internal integrity
  • Image intensifier removed and inspected for any imperfections and grid integrity
  • Camera removed and cleaned
  • Wheels removed, inspected, cleaned and or replaced from main chassis
  • Chassis structurally repaired
  • Vertical lift assembly is tested for drift and adjusted or replaced
  • All bearings are inspected, cleaned and relubricated, or replaced
  • Control panel display is tested and replaced
  • All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
  • All locks and brake assemblies are inspected, cleaned and repaired or replaced
  • Steering is tested and adjusted
  • Foot switch and hand switch provided on each unit
  • Operator and service manuals are provided
  • Re-assembly
  • All components are staged and sent to industrial bay for rebuilding
  • “C”, Image tube and camera, Vertical column, main control and power cable installed
  • New batteries are installed
  • New CRT monitors installed in monitor cart
  • Monitor cart rebuilt
  • When C-arm is completely rebuilt mechanically system is sent for calibration and final testing
  • Testing/calibration
  • Image intensifier and X-ray tube are tested for stability and balance
  • Image intensifier is tested for resolution and gain to OEM specifications
  • X-ray tube bearings are tested for noise and coast time
  • X-ray tube filaments and stator windings are tested
  • X-ray tube radiation output is verified to OEM specifications
  • X-ray generator high voltage tested
  • Maximum dose rate is set in compliance with FDA guidelines
  • CCD camera calibrated to OEM specifications
  • X-ray beam is aligned for every field size
  • C-arm is tested for leakage
  • Monitors are aligned, resized, centered and focused
  • Monitors are adjusted for brightness and linearity
  • Video system is optimized for gray scale and resolution
  • Technical Specifications Generator
  • High frequency generator 4.0 kW full-wave
  • Up to 120 kVp and 75 mA for radiographic exposures
  • Continuous fluro-mode up to 5 mA
  • One shot frame integration (low, medium, high)
  • Full power from standard 110 V
  • 15 A or 220 V 8 A outlet
  • Patented energy buffer design
  • X-Ray Tube
  • Rotating anode X-ray tube
  • Focal spots: 0.3 - 0.6 mm
  • 300,000 HU Anode heat capacity
  • Physical Specifications
  • Free Space in Arc: 31"
  • Depth of Arc: 26"
  • Arc orbital movement: 115 degrees
  • Left/Right wig-wag scan: +/- 11 degrees
  • Vertical Travel: 18" motorized
  • Horizontal travel: 8"
  • L-arm rotation: +/- 180 degrees
  • C-arm Dimensions
  • Length: 78.5"
  • Height: 68.25"
  • Width: 33"
  • Workstation Dimensions
  • Width: 27.25"
  • Height: 68.25"
  • Depth: 27.25"

Close Window