OEC 9600
Remanufacturing Process
Huestis Medical remanufactured OEC 9600 C-arms offer excellent imaging results at
a remarkable price. Even more remarkable is our ability to deliver performance and
safety features guaranteed to meet or exceed original manufacturer’s specification.
Thoroughly remanufactured, our OEC 9600 C-arms are offered with new and upgraded
imaging components. We disassemble, mechanically restore, refinish, calibrate and
pre-stage each C-arm prior to delivery.
Process / Features
- Mechanical repairs/cosmetic maintenance
Computer controlled ABS, mA, kVp and camera gain
All defective parts and covers are repaired or replaced
C-arm is completely disassembled, disinfected and repainted
“C” is removed from main frame assembly
Horizontal arm removed, inspected and repaired as necessary
Vertical column removed and bearings removed,
inspected and replaced
X-ray tube removed and inspected for artifacts and
internal integrity
Image intensifier removed and inspected for any imperfections
and grid integrity
Camera removed and cleaned
Wheels removed, inspected, cleaned and or replaced from
main chassis
Chassis structurally repaired
Vertical lift assembly is tested for drift and adjusted
or replaced
All bearings are inspected, cleaned and relubricated,
or replaced
Control panel display is tested and replaced
All wheels are inspected and lubricated and/or replaced
as needed, with new cable pushers installed
All locks and brake assemblies are inspected, cleaned
and repaired or replaced
Steering is tested and adjusted
Foot switch and hand switch provided on each unit
Operator and service manuals are provided
- Re-assembly
All components are staged and sent to industrial bay
for rebuilding
“C”, Image tube and camera, Vertical column, main control
and power cable installed
New batteries are installed
New CRT monitors installed in monitor cart
Monitor cart rebuilt
When C-arm is completely rebuilt mechanically system
is sent for calibration and final testing
Testing/calibration
Image intensifier and X-ray tube are tested for
stability and balance
Image intensifier is tested for resolution and
gain to OEM specifications
X-ray tube bearings are tested for noise and coast time
X-ray tube filaments and stator windings are tested
X-ray tube radiation output is verified to OEM specifications
X-ray generator high voltage tested
Maximum dose rate is set in compliance with FDA guidelines
CCD camera calibrated to OEM specifications
X-ray beam is aligned for every field size
C-arm is tested for leakage
Monitors are aligned, resized, centered and focused
Monitors are adjusted for brightness and linearity
Video system is optimized for gray scale and resolution
- Technical Specifications Generator
High frequency generator 4.0 kW full-wave
Up to 120 kVp and 75 mA for radiographic exposures
Continuous fluro-mode up to 5 mA
One shot frame integration (low, medium, high)
Full power from standard 110 V
15 A or 220 V 8 A outlet
Patented energy buffer design
- X-Ray Tube
Rotating anode X-ray tube
Focal spots: 0.3 - 0.6 mm
300,000 HU Anode heat capacity
- Physical Specifications
Free Space in Arc: 31"
Depth of Arc: 26"
Arc orbital movement: 115 degrees
Left/Right wig-wag scan: +/- 11 degrees
Vertical Travel: 18" motorized
Horizontal travel: 8"
L-arm rotation: +/- 180 degrees
- C-arm Dimensions
Length: 78.5"
Height: 68.25"
Width: 33"
- Workstation Dimensions
Width: 27.25"
Height: 68.25"
Depth: 27.25"
|