OEC 9400
Remanufactured C-Arm
Remanufactured OEC 9400 C-arms from Huestis Medical offer
excellent imaging results at a remarkable price. Even more
remarkable is our ability to deliver performance and safety
features guaranteed to meet or exceed original manufacturer’s
specifications.
Thoroughly remanufactured, our OEC 9400 C-arms are restored to
like-new condition through multi-step processes. We disassemble,
mechanically and electronically restore, refinish, calibrate, and
perform final testing and QA on each C-arm. Our full line of OEC
C-arms offers imaging excellence and reliability, at an affordable price.
X-Ray System
- Generator
- High frequency, 2.5 kHz
- 7.5 kW full-wave
- Up to 120 kVp
- Up to 100 mA for radiographic film exposures
- Boosted fluoro and Pulsed fluoro capability
- Full power from standard outlet 120V/15A
- Patented energy buffer design
- X-Ray Tube
- Rotating anode X-ray tube
- Focal spots: 0.3 mm and 1.0 mm
- Focal spot kW rating: 0.3 mm = 5.3 kW and 1.0 mm = 41.0 kW
- Anode heat capacity: 300,000 H.U.
- Anode cooling rate: 70,000 H.U./min
- Housing heat capacity: 1,250,000 H.U.
- Anode diameter: 3"
- Anode angle: 10°
- Collimator
- Two sets of opposing shutters
- Continuously variable and rotatable
- All functions remotely controlled from the C-Arm control panel
and optional hand held remote
- Fluoroscopy Mode
- Focal spot: 0.3 mm
- kVp range: 40 – 120 kVp
- Maximum ripple: typically 1% at 120 kVp/5 mA
- mA range: 0.2 – 5.0 mA normal mode, 1.0 – 20 mA boost mode
- Auto and manual modes
- Continuous, one-shot or pulsed operation
- ABS varies mA, kVp and camera gain
- User selectable ABS tables for orthopedics, chest and low dose
applications
- Pulsed Fluoroscopy Mode
- Focal spot: 0.3 mm
- kVp range: 40 – 120 kVp
- mA range: Up to 60 mA
- Pulse rate: 1, 2, 4, or 8 pulses per second
- Pulse width: 30 or 50 milliseconds
- Camera operates in progressive scan mode
- Computer controlled camera iris, mA, kVp and camera gain
- Radiographic Mode
- Focal spot: 0.3 mm or 1.0 mm
- Focal spot automatically selected
- mAs range: 1 – 300 mAs
- kVp range: 50 – 120 kVp
- Cassette holder (optional) 10" x 12" (24 cm x 30 cm),
9" Image Intensifier
Video Imaging System
- Image Intensifier
- Tri-mode 9"/6"/4.5" image intensifier
- Central resolution (typical):
4.5": 54 lp/cm
6": 48 lp/cm
9": 41 lp/cm
- Peripheral resolution at 70% radius (typical):
4.5": 50 lp/cm
6": 46 lp/cm
9": 40 lp/cm
- Contrast Ratio: 30:1 (typical)
- DQE: 65% (typical at 1.7 Hz)
- TV Camera
- 1 in Vidicon (high sensitivity)
- Progressive scan operation
- 360° motorized rotation
- On-screen orientation indicator
(real-time feedback without fluoro)
- Left-right image reversal
- Top-bottom image reversal
- Negative mode
- Bandwidth: 15 MHz
- Video signal: Standard RS 170 A, 60 Hz, 525 line
- Aspect ratio: 4:3
- Computer controlled features:
- Dark current compensation
- Gain
- Blanking
- Camera iris
- Video Monitors
- Dual 17 in ultrabrite anti-glare monitors
- 525 lines, 20 MHz
- Ambient room light sensor
- Remote brightness/contrast controls
Image Processing
- Expanded Surgical Platform (ESP)
- 60 image storage with last image hold
- 640 x 512 x 10 bit
- Frame averaging noise reduction (low, medium, high)
- One-shot frame integration (low, medium, high)
- MARS (motion artifact reduction system)
- Variable edge enhancement (track ball)
- Digital window/level
- Real-time auto-histogramming (auto window/level)
- Variable zoom and roam (track ball)
- Patient annotation keyboard
- Anatomical markers
- On-line help menus
- Chole mode (up to 20 mA)
- 4 fps digital disc (optional)
- Vascular Module
- Includes all Expanded Surgical Platform (ESP) features plus:
- Real-time subtraction
- Roadmapping
- Peak opacification
- Re-registration (track ball)
- Variable landmarking (track ball)
- Mask save/recall
- 4 fps Digital Disc (optional)
- Record rate: 1, 2 or 4 Frames/Sec (fps)
- Record time: 20, 10 or 5 minutes
- Play mode: 1, 2 or 4 Frames/Sec (fps)
- Control: Instant image access (track ball)
- Synchronized to pulse with generator,
T.V. camera and image processor
Hardcopy Options
- Radiographic film capability
- Thermal printer
System Control
- Entire system is computer controlled
- Software upgradable
- Main control panel pivots right-to-left
for user convenience
- Multi-function infrared remote control (optional)
- Hand held X-ray remote control (optional)
- Multi-function footswitch
Mechanical
- Source to image distance: 36.18" (919 mm)
- Free space in arc: 27.2" (691 mm)
- Depth of arc: 23.3" (592 mm)
- Arc orbital movement: 111°
- Left/right wig-wag scan: ±11°
- Vertical travel: 18" motorized (457 mm)
- Horizontal travel: 8" (203 mm)
- L-Arm rotation: ±185° motorized
- Reversible C-Arm: 180° manual flip-flop
C-Arm Dimensions
- In compacted position:
- Length: 74.5" (1892 mm)
- Width: 34" (864 mm)
- Height: 65.8" (1671 mm)
Workstation Dimensions
- Depth: 27" (686 mm)
- Width: 32" (813 mm)
- Height: 65" (1651 mm)
Electrical
- Input power (50 Hz or 60 Hz):
115 V, 15 A
230 V, 12 A
100 V, 20 A
|